Protocol Analysis
R&S
®
Scope Rider RTH
214User Manual 1326.1578.02 ─ 12
"PetrolLevel", starts at bit #24, has a length of 8 bit, and the unit is liter. The signal
"IdleRunning" is state-encoded. It has only one bit. The binary value 0 means "Run-
ning", and the binary value 1 means "Idle".
Example: CAN DBC file section
BO_ 2166573756 EngineData: 8 Engine
SG_ PetrolLevel : 24|8@1+ (1,0) [0|255] "l" ...
SG_ EngPower : 48|16@1+ (0.01,0) [0|350] "kW" ...
SG_ EngForce : 32|10@1+ (1,0) [0|1000] "N" ...
SG_ IdleRunning : 23|1@1+ (1,0) [0|1] "" ...
SG_ EngTemp : 16|7@1+ (2,-50) [-50|150] "degC" ....
SG_ EngSpeed : 0|13@1+ (1,0) [0|8000] "rpm" ...
....
VAL_ 2166573756 IdleRunning 0 "Running" 1 "Idle" ;
If a DBC file is loaded, the bus menu provides an additional entry: Display Labels.
Figure 9-27: Display of an CAN DBC file
CAN and CAN FD (Options R&S RTH-K3, R&S RTH-K9)