Remote Control Commands
R&S
®
Scope Rider RTH
304User Manual 1326.1578.02 ─ 12
ECL
-1.3 V
CMOS
2.5 V
GND
0 V (for CAN channels, requires option R&S RTH-K3)
CAN
2 V (for CAN channels, requires option R&S RTH-K3)
LIN7vsupply | LIN12vsupply | LIN18vsupply
7 V / 12 V / 18 V (for LIN channels, requires option R&S RTH-
K3)
USER
Set the value with CHANnel<m>:THReshold:USER.
*RST: 
TTL
CHANnel<m>:THReshold:USER <ThresholdValue>
Set an individual threshold value if CHANnel<m>:THReshold:TECHnology is set to
USER.
Suffix: 
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<ThresholdValue> Range:  -400  to  400
Increment:  1E-3
*RST:  1.4
Default unit: V
CHANnel<m>:THReshold:THReshold?
Returns the threshold value.
Suffix: 
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Return values: 
<Level> Range:  -10  to  10
Increment:  1E-3
*RST:  0
Default unit: V
Usage:  Query only
CHANnel<m>:THReshold:FINDlevel
The instrument sets the threshold for the selected channel.
Suffix: 
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Waveform Setup