TIMebase:REFerence <ReferencePoint>
Defines the time reference point in the diagram.
Parameters:
<ReferencePoint> Position of the reference point in percent of the screen width.
Available values are:
10: on the left side of the screen
50: in the middle of the screen
90: on the right side
*RST:  50
Default unit: %
15.3.4  Acquisition Control
RUN............................................................................................................................. 306
STOP........................................................................................................................... 306
ACQuire:MODE............................................................................................................. 306
ACQuire:AVERage:COUNt..............................................................................................307
ACQuire:ARESet:IMMediate........................................................................................... 307
ACQuire:POINts:PRESelect............................................................................................307
ACQuire:WAVeformupd.................................................................................................. 308
ACQuire:POINts:ARATe?................................................................................................308
ACQuire:POINts[:VALue]?...............................................................................................308
ACQuire:RESolution?.....................................................................................................308
RUN
Starts the continuous acquisition.
Usage: 
Event
STOP
Stops the running acquistion.
Usage: 
Event
ACQuire:MODE <AcquisitionMode>
Defines how the waveform is built from the captured samples.
Parameters:
<AcquisitionMode> SAMPle | PDETect | HRESolution | AVERage | ENVelope
SAMPle
One of n samples in a sample interval is recorded as waveform
point, the other samples are discarded.