Protocol Analysis
R&S
®
Scope Rider RTH
209User Manual 1326.1578.02 ─ 12
CAN Trigger................................................................................................................ 209
Frame type.................................................................................................................. 210
ID type.........................................................................................................................210
Identifier Pattern..........................................................................................................210
Identifier Relation.........................................................................................................211
FDF Bit........................................................................................................................ 211
BRS Bit........................................................................................................................211
ESI Bit..........................................................................................................................211
Data Pattern................................................................................................................ 211
Data Relation...............................................................................................................211
Byte Offset...................................................................................................................212
Address from Label..................................................................................................... 212
Error conditions: CRC, Bit stuffing, Form, Ack, SC..................................................... 212
CAN Trigger
Depending on the selected CAN trigger type, different additional parameters are avail-
able.
"Start of Frame"
Triggers on the stop bit of the sync field.
"End of frame"
Triggers after a wakeup frame.
"Frame type"
Triggers on a specified frame type (data, remote, error, or overload).
For data and remote frames, also the identifier format is considered.
"Identifier"
Sets the trigger to a specific identifier or an identifier range. Only the
6-bit identifier without parity bits is considered, not the protected iden-
tifier.
"Identifier + Data"
Sets the trigger to a combination of identifier and data condition. The
instrument triggers at the end of the last byte of the specified data
pattern.
CAN and CAN FD (Options R&S RTH-K3, R&S RTH-K9)