Remote Control Commands
R&S
®
Scope Rider RTH
305User Manual 1326.1578.02 ─ 12
Usage: Event
15.3.3 Horizontal Setup
TIMebase:SCALe...........................................................................................................305
TIMebase:RANGe..........................................................................................................305
TIMebase:HORizontal:POSition.......................................................................................305
TIMebase:REFerence.....................................................................................................306
TIMebase:SCALe <Scale>
Sets the time scale of the horizontal axis for all signals.
Parameters:
<Scale> Range: 1E-9 to 500
Increment: Steps 1, 2, 5 (1, 2, 5, 10, 20, 50, 100, 200, 500...)
*RST: 100E-9
Default unit: s/div
TIMebase:RANGe <AcquisitionTime>
Sets the acquisition time, the timerange across the 10 horizontal divisions of the dia-
gram. Use the command alternatively to TIMebase:SCALe.
Parameters:
<AcquisitionTime> Range: 10E-9 to 5000
Increment: Steps 1, 2, 5 (1, 2, 5, 10, 20, 50, 100, 200, 500...)
*RST: 1E-6
Default unit: s
TIMebase:HORizontal:POSition <Position>
Sets the horizontal position of the trigger point in relation to the reference point.
See also: "Horizontal Position" on page 46
Parameters:
<Position> Range: Depends on the time scale, see table below.
Increment: 0.1
*RST: 0
Default unit: s
Time scale Min./max. horizontal position
1 ns/div to 100 µs/div ±2 s
200 µs/div to 2 s/div Time scale * 20000
5 s/div – 500 s/div 100000 s
Waveform Setup