Frequency Counter (R&S RTH-K33)
R&S
®
Scope Rider RTH
262User Manual 1326.1578.02 ─ 12
If the measured reference frequency deviates extremely from the specified fre-
quency, a warning is displayed.
11. Select "Restart" to reset the statistical values and start a new measurement with
the reference frequency.
The deviation of the nominal frequency to the measured frequency is internally
removed from the measured frequency. The display indicates the accurate fre-
quency of the input signal.
12. Optionally, tap the second counter tab again to display both counter results simulta-
neously.
13. To start and stop a counter measurement, press the [RUN STOP] key.
11.4 Counter Settings
Access: "Counter" menu
Additional settings for vertical scaling are available for Counter mode and directly
accessible via the "Counter" menu. For a description see:
●
"Channel Index" on page 41
●
"Probe Setting" on page 42
●
"Bandwidth" on page 42
Furthermore, the logger function can be used to store counter values, see Chapter 8,
"Data Logging", on page 155.
Counter Settings