Remote Control Commands
R&S
®
Scope Rider RTH
302User Manual 1326.1578.02 ─ 12
CHANnel<m>:POSition <Position>
Moves the selected signal up or down in the diagram. The position is a graphical set-
ting given in divisions, while the offset sets a voltage.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Position> Position value, given in divisions.
Range: -4 to 4
Increment: 0.5
*RST: 0
Default unit: div
CHANnel<m>:OFFSet <Offset>
Sets an offset voltage that is added to correct an offset-affected signal. The value is
included in measurement results. The signal is shifted in relation to the ground level by
the offset value. Negative offset values move the waveform down, positive values
move it up.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Offset> Range: -400 to 400
Increment: 0.5
*RST: 0
Default unit: V
CHANnel<m>:COUPling <Coupling>
Selects the connection of the indicated channel.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
Parameters:
<Coupling> DCLimit | ACLimit
DCLimit
The signal passes the input unchanged.
ACLimit
A high-pass filter removes the DC offset voltage from the input
signal if the DC component of a signal is of no interest.
*RST: DCLimit
CHANnel<m>:BANDwidth <BandwidthLimit>