Remote Control Commands
R&S
®
Scope Rider RTH
311User Manual 1326.1578.02 ─ 12
Suffix:
<m>
.
1..21
Indicates the trigger source:
1..4: analog channels 1 to 4
8..15: digital channels D0 to D7
all others: not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger:HOLDoff:MODE
<Mode>
Selects the method to define the holdoff.
Parameters:
<Mode> OFF | TIME | RANDom | EVENts
OFF
No holdoff
TIME
Defines the holdoff as a time period. The next trigger occurs only
after a time has passed, which is defined with TRIGger:
HOLDoff:TIME.
RANDom
Defines the holdoff as a random time limited by TRIGger:
HOLDoff:MIN and TRIGger:HOLDoff:MAX. For each acquisi-
tion, the instrument selects a new random holdoff time from the
specified range.
EVENts
Defines the holdoff as a number of trigger events, which is
defined with TRIGger:HOLDoff:EVENts.
*RST:
OFF
TRIGger:HOLDoff:TIME <Time>
Sets the time that has to pass at least until the next trigger occurs. The command
takes effect if TRIGger:MODE is set to TIME.
Parameters:
<Time> Range: 8E-9 to 10
Increment: 200E-6
*RST: 1E-3
Default unit: s
Waveform Setup