Remote Control Commands
R&S
®
Scope Rider RTH
413User Manual 1326.1578.02 ─ 12
SINGle
Single measurement; the continuous measurement is stopped
and the most recent value is preserved.
*RST: AUTO
15.10 Data Logging
Logger Settings.....................................................................................................413
Scaling.................................................................................................................. 416
Cursor in Logger Mode......................................................................................... 418
Zoom in Logger mode...........................................................................................420
Logger Statistics ...................................................................................................421
Logger Records.....................................................................................................424
15.10.1 Logger Settings
LOGGer:SOURce.......................................................................................................... 413
LOGGer:TIMebase:SRATe.............................................................................................. 413
LOGGer:TIMebase:SCALe..............................................................................................414
LOGGer:SLOT:CURRent................................................................................................ 414
LOGGer:SLOT:SLOT<m>:NAME.....................................................................................414
LOGGer:SLOT:SLOT<m>:HASData?............................................................................... 414
LOGGer:SLOT:SLOT<m>:STARt?................................................................................... 415
LOGGer:SLOT:LOAD..................................................................................................... 415
LOGGer:SLOT:CLEar.....................................................................................................415
LOGGer:ACLR.............................................................................................................. 415
LOGGer:SOURce <Source>
Sets the logger source.
Befor you can change the logger source, stop the measurement.
Parameters:
<Source> SCOPe | METer | COUNter
SCOPe: an active automatic measurement is required.
METer | COUNter: The selected measurement must be active,
but not running.
LOGGer:TIMebase:SRATe <NextSampleRate>
Sets the number of samples per second.
Parameters:
<NextSampleRate> SA1 | SA2 | SA5 | SA10
*RST: SA1