EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 415

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
415User Manual 1326.1578.02 ─ 12
Return values:
<HasData> 1 | 0
1: data available
0: slot is empty
*RST: 0
Usage: Query only
Firmware/Software: FW 1.60 and higher
LOGGer:SLOT:SLOT<m>:STARt? <Year>, <Month>, <Day>, <Hour>, <Min>, <Sec>
Returns the start time of the logger record. If the slot is empty, the command returns an
error.
Suffix:
<m>
.
1..10
Slot number
Query parameters:
<Year>, <Month>,
<Day>, <Hour>,
<Min>, <Sec>,
<ZSec>
Date and time of the slot start
Usage: Query only
Firmware/Software: FW 1.60 and higher
LOGGer:SLOT:LOAD
Loads the logger record of a slot, and activates the slot. The slot is defined using
LOGGer:SLOT:CURRent.
Usage:
Event
LOGGer:SLOT:CLEar
Deletes the log data of a slot. The slot is defined using LOGGer:SLOT:CURRent.
Only possible while logging is stopped.
Usage:
Event
LOGGer:ACLR
Deletes the log data of all slots. Only possible while logging is stopped.
Usage: