Remote Control Commands
R&S
®
Scope Rider RTH
321User Manual 1326.1578.02 ─ 12
TRIGger:PATTern:WIDTh:MINWidth <MinDuration>
Sets the lower time limit of the pattern match for range conditions WITHin and OUT-
Side.
Parameters:
<MinDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:PATTern:WIDTh:MAXWidth <MaxDuration>
Sets the upper time limit of the pattern match for range conditions WITHin and OUT-
Side.
Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.8 State Trigger (R&S RTH-K19)
See also Chapter 3.6.8, "State Trigger (R&S RTH-K19)", on page 64.
TRIGger:STATe:CHANnel<m>.........................................................................................321
TRIGger:STATe:COMBination..........................................................................................322
TRIGger:STATe:CSOurce[:VALue]................................................................................... 322
TRIGger:STATe:CSOurce:EDGE..................................................................................... 322
TRIGger:STATe:CHANnel<m> <State>
Sets the state of each input channel. The channel is specified by the channel suffix:
The logical combination of the channel states is defined by TRIGger:STATe:
COMBination.
Suffix:
<m>
.
1..4: analog channel 1 to 4
8..15: digital channels D0 to D7
Parameters:
<State> ONE | ZERO | DONTcare
ONE
The signal value is above the defined threshold.
ZERO
The signal value is below the defined threshold.
DONTcare
The signal state does not matter.
*RST: DONTcare