Remote Control Commands
R&S
®
Scope Rider RTH
307User Manual 1326.1578.02 ─ 12
PDETect
The minimum and the maximum of n samples are recorded as
waveform points, the other samples are discarded.
HRESolution
The average of n captured sample points is recorded as one
waveform point.
AVERage
The average is calculated from the data of the current acquisi-
tion and a number of acquisitions before. The number of acquisi-
tions for average calculation is defined with ACQuire:
AVERage:COUNt.
ENVelope
The minimum and maximum values in an sample interval over a
number of acquisitions are saved. The most extreme values of
all acquisitions build the envelope.
*RST:
SAMPle
ACQuire:AVERage:COUNt
<NoOfAvgs>
Sets the number of waveforms used to calculate the average waveform.
Parameters:
<NoOfAvgs> Range: 2 to 8192
Increment:
2
N
(N = 1 .. 13)
*RST: 2
ACQuire:ARESet:IMMediate
Restarts the envelope and average calculation.
Usage:
Event
ACQuire:POINts:PRESelect <RecLength>
Sets the record length.
Parameters:
<RecLength> MAX | MIDDle | MIN
MAX
Sets the maximum possible record length.
MIDDle
Limits the record length to 12.5 kSample.
MIN
Limits the record length to 1.25 kSample.
*RST: MAX
The actual record length can differ from these limits, see "Act. Record Len. C1 - C4"
on page 48.