Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.15 Window Trigger (R&S RTH-K19)
TRIGger:LEVel<m>:WINDow:LOWer............................................................................... 332
TRIGger:LEVel<m>:WINDow:UPPer................................................................................332
TRIGger:WINDow:TIME..................................................................................................332
TRIGger:WINDow:RANGe.............................................................................................. 333
TRIGger:WINDow:WIDTh............................................................................................... 333
TRIGger:WINDow:DELTa................................................................................................333
TRIGger:WINDow:MINWidth........................................................................................... 333
TRIGger:WINDow:MAXWidth..........................................................................................334
TRIGger:LEVel<m>:WINDow:LOWer <Level>
TRIGger:LEVel<m>:WINDow:UPPer <Level>
Set the upper and lower voltage thresholds for the window trigger, respectively. The
trigger levels are the vertical window limits.
Suffix:
<m>
.
1..4
Indicates the trigger source: analog channel 1 to 4
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
TRIGger:WINDow:TIME <Condition>
Selects how the time limit of the window is defined.
Parameters:
<Condition> LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
LONGer | SHORter
Triggers if the signal crosses the upper or lower level after/
before the time "Width" defined by TRIGger:WINDow:WIDTh
EQUal | NEQual
Triggers if the signal stays inside or outside the vertical window
limits for a time equal/uneuqal to "Width" "±Tolerance" defined
by TRIGger:WINDow:WIDTh and TRIGger:WINDow:DELTa.
WITHin
Triggers if the signal stays inside or outside the vertical window
limits for a time "Min Width" AND "Max Width".