Remote Control Commands
R&S
®
Scope Rider RTH
333User Manual 1326.1578.02 ─ 12
OUTSide
Triggers if the signal stays inside or outside the vertical window
limits for a time < "Min Width" OR > "Max Width".
*RST: LONGer
TRIGger:WINDow:RANGe <LevelRangeMode>
Selects how the signal run is compared with the window:
Parameters:
<LevelRangeMode> ENTer | EXIT | WITHin | OUTSide
ENTer | EXIT
Triggers when the signal crosses the upper or lower level and
thus enters/leaves the window made up of these two levels
which are defined by TRIGger:LEVel<m>:WINDow:UPPer
and TRIGger:LEVel<m>:WINDow:LOWer
WITHin | OUTSide
Triggers if the signal stays between/above the upper and lower
level for a specified time. The time is defined by TRIGger:
WINDow:TIME
*RST: ENTer
TRIGger:WINDow:WIDTh <Duration>
Sets the width for comparison ranges LONGer, SHORter, EQUal, NEQual.
See TRIGger:WINDow:TIME on page 332.
Parameters:
<Duration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:WINDow:DELTa <Tolerance>
Sets a range Δt to the specified width, which is defined using TRIGger:WINDow:
WIDTh.
Parameters:
<Tolerance> Range: 0 to 5000
Increment: 500E-12
*RST: 0
Default unit: s
TRIGger:WINDow:MINWidth <MinDuration>
Sets the lower time limit of the stay inside or outside the window (comparisons WITHin
and OUTSide).
Waveform Setup