EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - General Instrument Setup

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
501User Manual 1326.1578.02 ─ 12
Setting parameters:
<FileSource> String parameter
Path and name of the file to be moved.
<FileDestination> String parameter
Path and name of the new file.
Example:
MMEM:MOVE "/media/SD/Rohde-Schwarz/RTH/SaveSets/Settings1.xml",
"/media/USB1/SaveSets/Settings1.xml"
Usage: Setting only
MMEMory:DELete
<FileSource>
Removes a file from the specified directory.
Setting parameters:
<FileSource> String parameter
File name and path of the file to be removed. If the path is omit-
ted, the specified file is deleted in the current directory. Filters
are not allowed.
Example:
MMEM:DELete "/media/USB1/SaveSets/Settings1.xml"
Usage: Setting only
15.14 General Instrument Setup
Date and Time.......................................................................................................501
Display Settings.................................................................................................... 502
15.14.1 Date and Time
SYSTem:DATE [<Year>], [<Month>], [<Day>]
SYSTem:DATE? [<Year>], [<Month>]
Sets the date on the instrument.
Parameters:
<Day> Range: 1 to 31
Increment: 1
*RST: 1
Parameters for setting and query:
<Year> Range: 2012 to 2099
Increment: 1
*RST: 2012
General Instrument Setup