Remote Control Commands
R&S
®
Scope Rider RTH
502User Manual 1326.1578.02 ─ 12
<Month> Range: 1 to 12
Increment: 1
*RST: 1
SYSTem:TIME [<Hours>], [<Minutes>], [<Seconds>]
SYSTem:TIME? [<Hours>], [<Minutes>]
Sets the time on the instrument.
Parameters:
<Seconds> Range: 0 to 59
Increment: 1
*RST: 1
Parameters for setting and query:
<Hours> Range: 0 to 24
Increment: 1
*RST: 1
<Minutes> Range: 0 to 59
Increment: 1
*RST: 1
15.14.2 Display Settings
DISPlay:PERSistence[:TYPE]......................................................................................... 502
DISPlay:PERSistence:TIME............................................................................................503
DISPlay:CONTrast......................................................................................................... 503
DISPlay:MOUS..............................................................................................................503
DISPlay:LCD................................................................................................................. 503
DISPlay:PERSistence[:TYPE] <PersistenceType>
Defines how long every new data point remains on the screen.
Parameters:
<PersistenceType> OFF | OFFM | TIME | INF
OFF
Deactivates persistence and displays the last acquisition.
OFFM
Deactivates persistence and displays multiple acquistions.
TIME
Data points remain on the screen for the duration defined with
DISPlay:PERSistence:TIME.
INF
Data points remain on the screen infinitely until persistence is
set to OFF.
*RST:
OFF