TRIGger:WIDTh:MAX <MaxDuration>
Sets the upper limit for the pulse width if TRIGger:WIDTh:RANGe is set to WITHin or
OUTSide.
Parameters:
<MaxDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:WIDTh:MIN <MinDuration>
Sets the lower limit for the pulse width if TRIGger:WIDTh:RANGe is set to WITHin or
OUTSide.
Parameters:
<MinDuration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
15.3.5.5 Video/TV Trigger
The standards PAL, PAL-M, NTSC and SECAM are available in the instrument firm-
ware. All other standards and custom signals require the advanced trigger option
(R&S RTH-K19).
TRIGger:TV:STANdard................................................................................................... 315
TRIGger:TV:POLarity..................................................................................................... 316
TRIGger:TV:MODE........................................................................................................ 316
TRIGger:TV:LINE...........................................................................................................317
TRIGger:TV:LFIeld......................................................................................................... 317
TRIGger:TV:CUSTom:STYPe..........................................................................................317
TRIGger:TV:CUSTom:LDURation.................................................................................... 317
TRIGger:TV:CUSTom:SDURation....................................................................................317
TRIGger:TV:CUSTom:SCANmode................................................................................... 318
TRIGger:TV:STANdard <Standard>
Selects the TV standard or CUSTom for user-defined signals.
Parameters:
<Standard> PAL | PALM | NTSC | SECam | P480L60HZ | P576L50HZ |
P720L30HZ | P720L50HZ | P720L60HZ | I1080L50HZ |
I1080L60HZ | P1080L24HZ | P1080L24HZSF | P1080L25HZ |
P1080L30HZ | P1080L50HZ | P1080L60HZ | CUSTom
PAL | PALM | NTSC | SECam
Standards delivered with the instrument firmware.