Remote Control Commands
R&S
®
Scope Rider RTH
314User Manual 1326.1578.02 ─ 12
Positive going pulses. The width is defined from the rising to the falling edge.
Negative going pulses. The width is defined from the falling to the rising edge.
Both positive and negative going pulses
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:WIDTh:RANGe <Condition>
Defines how the measured pulse width is compared with the given limit(s).
Parameters:
<Condition> LONGer | SHORter | EQUal | NEQual | WITHin | OUTSide
LONGer | SHORter
Triggers on pulses shorter or longer than a width set using
TRIGger:WIDTh:WIDTh.
EQUal | NEQual
Triggers on pulses equal or not equal a given width that is set
using TRIGger:WIDTh:WIDTh. In addition, a tolerance can be
set around the specified width using TRIGger:WIDTh:DELTa.
WITHin | OUTSide
Triggers on pulses inside or outside a given range. The range is
set using TRIGger:WIDTh:MIN and TRIGger:WIDTh:MAX.
*RST:
LONGer
TRIGger:WIDTh:WIDTh <Duration>
Sets the width for comparison ranges EQUal, UNEQual, SHORter, and LONGer.
See TRIGger:WIDTh:RANGe on page 314
Parameters:
<Duration> Range: 800E-12 to 10000
Increment: 100E-9
*RST: 5E-9
Default unit: s
TRIGger:WIDTh:DELTa <Tolerance>
Sets a range Δt to the specified width, which is defined using TRIGger:WIDTh:WIDTh
.
Parameters:
<Tolerance> Range: 0 to 5000
Increment: 500E-12
*RST: 0
Default unit: s