EasyManua.ls Logo

Rohde & Schwarz Scope Rider RTH1004 - Page 414

Rohde & Schwarz Scope Rider RTH1004
530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
Scope Rider RTH
414User Manual 1326.1578.02 ─ 12
LOGGer:TIMebase:SCALe <NextHorizScale>
Selects the horizontal scale of the logged data.
Parameters:
<NextHorizScale> AUTO | S1 | S2 | S4 | S5 | S10 | S20 | S40 | M1 | M2 | M4 | M5 |
M10 | M20 | M40 | H1 | H2 | H4 | H5 | H10 | H20 | D1 | D2 | D4
S1 | S2 | S4 | S5 | S10 | S20 | S40
Seconds per division
M1 | M2 | M4 | M5 | M10 | M20 | M40
Minutes per division
H1 | H2 | H4 | H5 | H10 | H20
Hours per division
D1 | D2 | D4
Days per division
*RST: AUTO
LOGGer:SLOT:CURRent
<SelectedSlot>
Selects one of the 10 memory slots to store the data during the logging. It is possible to
change the slot during recording. This command shows the slot number that is
selected in the menu.
Parameters:
<SelectedSlot> SLOT1 | SLOT2 | SLOT3 | SLOT4 | SLOT5 | SLOT6 | SLOT7 |
SLOT8 | SLOT9 | SLOT10
*RST: SLOT1
LOGGer:SLOT:SLOT<m>:NAME <Name>
Sets an optional slot name to describe the logged data.
Suffix:
<m>
.
1..10
Slot number
Parameters:
<Name> String parameter, name of the slot. The maximum length of the
name is 20 characters.
Firmware/Software: FW 1.60 and higher
LOGGer:SLOT:SLOT<m>:HASData?
Queries if the slot has data.
Suffix:
<m>
.
1..10
Slot number
Data Logging

Table of Contents