Remote Control Commands
R&S
®
Scope Rider RTH
420User Manual 1326.1578.02 ─ 12
Suffix:
<n>
.
1..4
Specifies the measurement. You can log the results of up to four
active measurements.
Return values:
<ResultDelta> Absolute value of the result difference.
Usage: Query only
LOGGer:CURSor<m>:POSition
Returns or sets the timestamp (absolute time) of the specified cursor line (t1 and t2 on
the result display).
Suffix:
<m>
.
1..2
Specifies the cursor line.
Parameters:
<Time> List of values:
<Year>,<Month>,<Day>,<Hour>,<Minute>,<Second>,
<TenthsOfSecond>
Year, month, day, hour, minute, second, tenth of a second, for
example 2015,10,29,16,10,22,2
LOGGer:CURSor:TDELta?
Returns the time difference of the cursor lines (t1 and t2 on the result display).
Return values:
<DeltaTime> List of values:
<Day>;<Hour>;<Minute>;<Second>;<TenthsOfSecond>
Days, hours, minutes, seconds, tenth of a second, for example
1,9,10,22,5
Usage: Query only
15.10.4 Zoom in Logger mode
LOGGer:ZOOM:ENABle................................................................................................. 420
LOGGer:ZOOM:SCALe.................................................................................................. 421
LOGGer:ZOOM:POSition................................................................................................421
LOGGer:ZOOM:ENABle <Zoom Enabled>
Enables or disables the logger zoom.
Parameters:
<Zoom Enabled> ON | OFF
*RST: OFF