Protocol Analysis
R&S
®
Scope Rider RTH
250User Manual 1326.1578.02 ─ 12
Figure 9-45: Table display of a decoded SENT signal, triggered on pulse period error
Table 9-13: Content of the protocol table for decoded SENT signal
Column Description
# Frame index
Frame start Time of frame start
Status Status value
ID [hex] Identifier value, hexadecimal value
Values 8 bit [format] Values of data bytes. The data format is selected in the "Bus" menu.
CRC [bin] CRC sequence value
State Overall state of the frame.
"Incomplete" indicates that the frame is not contained completely in the
acquisition. Change the time scale, or move the reference point to the left to
get a longer acquisition.
Remote control commands are described in Chapter 15.11.7.3, "SENT Decode
Results", on page 480.