Remote Control Commands
R&S
®
Scope Rider RTH
383User Manual 1326.1578.02 ─ 12
Parameters:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 A|10 A|100 A|
1000 A.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range: 1 to 1000
*RST: 1
Default unit: A
METer:SENSe:FREQuency:VOLTage:RANGe:UPPer
<Range>
Sets a fixed voltage range for frequency measurements and turns off auto ranging.
Parameters:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 V|10 V|100 V|
1 kV.
You can enter any value between the minimum and maximum
value. The instrument uses the next suitable range.
Range: 1 to 1000
*RST: 1
Default unit: V
METer:SENSe:RESistance:RANGe:UPPer <Range>
Sets a fixed range for resistance measurements and turns off auto ranging.
Parameters:
<Range> <numeric value> | MIN | MAX | DEF
<numeric value>
The instrument has fixed measurement ranges: 1 k|10 k|
100 k|1 M|10 M|100 M.
You can enter any value between the minimum and maximum
value. The instrument sets the next suitable range.
Range: 1 kOhm to 100 MOhm
*RST: 1 kOhm
METer:SENSe:VOLTage:AC:RANGe:UPPer <Range>
METer:SENSe:VOLTage:DC:RANGe:UPPer <Range>
Sets a fixed voltage range and turns off auto ranging.
Digital Multimeter (R&S