Remote Control Commands
R&S
®
Scope Rider RTH
491User Manual 1326.1578.02 ─ 12
Parameters:
<Source> C1 | C2 | C3 | C4 | M1 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7
D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7: All active logic channels
are saved, not matter which Dx is specified in the command.
*RST: C1
EXPort:WAVeform:MULTichannel <MultiChExport>
If ON, all active waveforms (analog, digital and math) are exported.
If OFF, select the waveform for export with EXPort:WAVeform:SOURce.
Parameters:
<MultiChExport> ON | OFF
ON = 1, and OFF = 0
*RST: ON
Example: See Example "Saving all active waveforms with time values"
on page 490.
EXPort:WAVeform:INCXvalues <IncHorValues>
Includes horizontal values in the export data (time values).
Parameters:
<IncHorValues> ON | OFF
ON = 1, and OFF = 0
*RST: OFF
Example: See Example "Saving a single waveform with time values"
on page 489.
EXPort:WAVeform:DLOGging <DataLogging>
The command requires option R&S RTH-K15. History must be enabled before using
the command (CHAN:HIST:STAT ON).
The command includes the history data in the export files. If acquisition is running, the
command stops the acquisition.
History data is always saved in a zip file (compressed csv), see also Chapter 4.7.4,
"Exporting History Data", on page 100.
You can select a range of history segments with CHANnel<m>:HISTory:STARt and
CHANnel<m>:HISTory:STOP.
Parameters:
<DataLogging> ON | OFF
ON = 1, and OFF = 0
*RST: OFF