General Instrument Settings R&S SMC100A
1411.4060.32 4.28 E-1
Recall - File
Loads the selected configuration.
If an instrument setting in which a sweep was activated is stored, the
sweep is started when the recall command is called.
If an instrument setting which accesses lists is stored, this list is also
loaded.
If the list has been deleted in the meantime, an error message
appears when the instrument setting is loaded. If the list has been
overwritten in the meantime, the new entries will be used.
Remote-control command:
MEMM:LOAD:STAT 2,"/var/user/testfile.savrcl"
*RCL 2
Recall Immediate x
Loads the selected configuration from one of the three intermediate
memories.
If an instrument setting in which a sweep was activated is stored, the
sweep is started when the recall command is called.
If an instrument setting which accesses lists is stored, this list is also
loaded.
If the list has been deleted in the meantime, an error message
appears when the instrument setting is loaded. If the list has been
overwritten in the meantime, the new entries will be used.
A message appears if no instrument configuration is stored in this
memory.
Remote-control command:
*RCL 1