R&S SMC100A Alphabetical List of SCPI-Error Messages
1411.4060.32 9.5 E-1
Command protected (-203)
The desired command could not be executed as it was protected with a password.
Example:
The command CAL:FREQ? might be protected with a password.
Remedy
Use command :SYST:PROT1 OFF, 123456 to enable the
command.
SCPI:
Execution error - sets bit 4 in the ESR register
Data out of range (-222)
A value of the transmitted command was outside the legal range.
Example:
Command SOUR:FREQ only permits entries in the range of min to
max frequency.
SCPI:
Execution error - sets bit 4 in the ESR register
Data type error (-104)
The command contains an invalid value indication.
Example:
ON is indicated instead of a numeric value for frequency setting -
:FREQ ON
SCPI:
Command error - sets bit 5 in the ESR register
Device-specific error (-300)
Device-specific error not defined in greater detail.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Exponent too large (-123)
The magnitude of the exponent is too large.
SCPI:
Command error - sets bit 5 in the ESR register