R&S SMC100A Alphabetical List of Device-Specific Error Messages
1411.4060.32 9.15 E-1
Invalid EEPROM data (203)
Reading a EEPROM is possible, however the data are inconsistent.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Invalid filename extension (464)
The file extension is not valid for the desired operation.
Example: The file extension for user correction files is *.uco. It is not
possible to enter another file extension when storing a user
correction file.
SCPI:
Device-specific error - sets bit 3 in the ESR register
No current list (241)
There is no list selected. To execute the desired operation a list has to be selected
in the related menu. If no list is available, a new list must be created.
Example:
The user correction mode is enable without a list being selected.
SCPI:
Device-specific error - sets bit 3 in the ESR register
This modulation forces other modulations off (140)
A modulation has been switched on which cannot be used at the same time as an
already active modulation. The previous modulation has been switched off.
Example:
Enabling FM modulation switches PM modulation off.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Unknown list type specified (242)
The list type selected is not valid for the desired operation
Example: The file extension for user correction files is *.uco. It is not
possible to enter another file extension when selecting a user
correction file.
SCPI:
Device-specific error - sets bit 3 in the ESR register