EasyManua.ls Logo

Rohde & Schwarz SMC100A - Output Subsystem

Rohde & Schwarz SMC100A
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Description R&S SMC100A
1411.4060.32 6.20 E-1
OUTPut Subsystem
The OUTPut system contains the commands which set the properties of the RF output connector. The
properties of the LF output connector are set in the SOURce:LFOutput system, see section
"SOURce:LFOutput Subsystem".
Command Parameters Default
unit
Remark
OUTPut:AFIXed:RANGe:LOWer?
dBm Query only
OUTPut:AFIXed:RANGe:UPPer?
dBm Query only
OUTPut:AMODe AUTO | FIXed
OUTPut:BLANk:LIST:STATe ON | OFF
OUTPut:IMPedance
Ohm Query only
OUTPut:PROTection:CLEar
No query
OUTPut:PROTection:TRIPped
Query only
OUTPut[:STATe] ON | OFF
OUTPut[:STATe]:PON OFF | UNCHanged
OUTPut:AMODe AUTO | NORMal
The command switches the mode of the attenuator at the RF output (A
ttenuator MODe).
Parameters: AUTO
The attenuator is switched automatically. The level settings are made in the
full range..
FIXed
The level settings are made without switching the attenuator. When this
operating mode is switched on, the attenuator is fixed in its current position
and the resulting variation range is defined.
Example:
POW:ALC ON
'activates automatic level control for RF output.
OUTP:AMOD FIX
'sets the fixed mode with uninterrupted level for the RF output'.
*RST value Resolution Options SCPI
AUTO - Device-specific

Table of Contents