R&S SMC100A Command Description
1411.4060.32 6.67 E-1
SOURce:PULM and PGEN Subsystem
The PULM subsystem contains the commands for checking the pulse modulation. The external signal is
input at the PULSE EXT connector. The connector can be used as trigger input for internal pulse
m
odulation. The polarity of the connector can be selected.
The PGEN Subsystem contains the command for activating/deactivating the output of the video/sync
signal at the PULSE VIDEO connector.
The keyword SOURce is optional with commands and can be omitted.
Command Parameters Default
unit
Remark
[SOURce:]PGEN:STATe ON | OFF
[SOURce:]PULM:DELay 20ns .. 1s
s
[SOURce:]PULM:DOUBle:DELay 20ns .. 1s
s
[SOURce:]PULM:DOUBle:STATe ON | OFF
[SOURce:]PULM:DOUBle:WIDTh 20ns .. 1s
[SOURce:]PULM:MODE DOUBle | SINGle
[SOURce:]PULM:PERiod 5Ts .. 85.s | 100ns .. 85 s
[SOURce:]PULM:POLarity NORMal | INVerted
[SOURce:]PULM:SOURce INT | EXT
[SOURce:]PULM:STATe ON | OFF
[SOURce:]PULM:TRIGger:EXTernal:SLOPe POSitive | NEGative
[SOURce:]PULM:TRIGger:EXTernal:GATE:POLarity NORMal | INVerted
[SOURce:]PULM:TRIGger:MODE AUTO | EXTernal | EGATe
[SOURce:]PULM:WIDTh 20ns .. 1s
s
[SOURce:]PGEN:STATe ON | OFF
The command switches ON/OFF the output of the video/sync signal at the PULSE VIDEO connector
at the rear of the instrument.
Example:
PGEN:STAT OFF
'deactivates output of video/sync signal
*RST value Resolution Options SCPI
OFF - Compliant