General Instrument Settings R&S SMC100A
1411.4060.32 4.10 E-1
Hardware Config... - Setup-System
In the Hardware Config… menu, the installed assemblies together with their variants and revision
states can be displayed for servicing purposes. The Hardware Config menu under System is opened
u
sing the
SETUP
key.
Section Counter in the upper part of the menu shows the Operation Hours and the number of power-
on (Power On Counter).
Remote-control command:
DIAG:INFO:OTIM? Response: "37"
DIAG:INFO:POC? Response: "147"
A table below lists the installed assemblies. It is divided into the sections Common Assembly and RF
Assembly and Baseband Assembly
Assembly
Assembly name
Part Number
Part Number of assembly
More…
Opens a dialog with additional information on serial number, revision and slot of
assembly and used bus by assembly (serial bus or PCI bus).
Remote-control command (see chapter 'Remote control Commands' for parameter name for modules):
DIAG:BGIN? "RFBO" Response: "1406.7007.00 02.00 123456,MOD-FPGA 20.20.00 "