General Instrument Settings R&S SMC100A
1411.4060.32 4.14 E-1
Test Point... - Setup-Test
The Test Point... menu provides access to the test points available in the instrument. Test Point is a
protected procedure, which can be accessed if protection level 1 is disabled. The protection dialog is
called in the setup menu (see Protection - Setup-Protection, page 4.21).
When activated, the voltage of the selected test point is measured and displayed. A detailed description
of the test points can be found in Chapter 3 of the Service Manual which is supplied with the instrument.
The Test Point... menu under System is opened using the
SETUP key.
State Test Point
Activates the measurement of the voltage at the selected test point.
Remote-control command:
DIAG1:MEAS:POIN? 'DIAG_NO_DP'
Response: 2
Select Test Point
Calls the submenu for selecting the test point. The currently selected
key is shown next to the key.
Remote-control commands:
DIAG1:POIN:CAT? (Command lists all test points)
DIAG1:POIN? 'DIAG_SMC_BAB_P7V'
With remote control, voltage measurement starts as soon as the test
point is selected.
Voltage Test Point
Displays the measured voltage measurement at the selected test
point.
Remote-control command:
DIAG1:MEAS:POIN? 'DIAG_SMC_BAB_P7V'
Response: 2