Command Description R&S SMC100A
1411.4060.32 6.58 E-1
SOURce:PHASe Subsystem
This subsystem contains the commands for adjusting the phase of the RF output signal relative to a
reference signal of the same frequency.
The keyword SOURce is optional with commands and can be omitted.
Command Parameters Default unit Remark
[SOURce:]PHASe -359.9 deg ... +359.9 deg RAD
[SOURce:]PHASe:REFerence
No query
[SOURce:]PHASe -359.9 deg ... +359.9 deg
The command specifies the phase variation relative to the current phase. The variation can be
specified in RADians.
Example:
PHAS 2DEG
'changes the phase by 2 degrees relative to the current phase.
PHAS:REF
'adopts the set phase as the current phase'.
*RST value Resolution Options SCPI
0.0 deg 0.1 deg Compliant
[SOURce:]PHASe:REFerence
The command adopts the phase set with SOURce:PHASe:ADJust as the current phase.
The command triggers an event and therefore has no *RST value and no query form.
Example:
PHAS 0.1RAD
'changes the phase by 0.1 rad relative to the current phase.
PHAS:REF
'adopts the set phase as the current phase'.
*RST value Resolution Options SCPI
- - Compliant