General Instrument Settings R&S SMC100A
1411.4060.32 4.20 E-1
Display Settings - Setup-Environment
The Display Settings... menu provides access to the settings for the display. It is opened using the
SETUP
key under Environment.
Screen Saver
Activates/deactivates the screen saver of the instrument display. If
activated, the display lamp is shut off when no entries via front panel,
external mouse or external keyboard are made for a period of time
(Wait Time). The screen saver prolongs the life time of the display
lamp.
This mode is recommended for preserving the display especially if the
instrument is exclusively operated via remote control.
Remote-control command:
DISP:PSAV:STAT ON
Wait Time
Enters the idle time that must elapse before the display lamp is shut
off when no entries are made.
Remote-control command:: DISP:PSAV:HOLD 10
Remote Channel Settings - Setup-Remote
The Remote GPIB/Ethernet... menu provides access to the GPIB and Ethernet settings. The Remote
GPIB... menu is opened using the
SETUP key under Remote.
GPIB channel address
Sets the IEC/IEEE-bus address of the instrument.
Remote-control command:
SYST:COMM:GPIB:ADDR 28