Internal Adjustments - Setup-System
The R&S SMC Signal Generator is extremely accurate thanks to the integrated procedures for
adjustments.
All internal adjustments for which no external measuring equipment is needed can be started in the
Internal Adjustments... menu. The adjustments with external measuring equipment are described in
the Service Manual (on CD ROM, supplied with the instrument).
Adjustment is recommended if the temperature range in which the instrument is operated changes, or
prior to all applications which require maximum level and frequency accuracy.
During adjustment a bar indicates the status of progress. If an error occurs, adjustment is terminated
and an error message is output in the info line.
NOTICE
Risk of invalid adjustment
In order to achieve correct adjustment of the instrument, make sure that the
instrument is warm before performing adjustments. The warm-up time is 30 minutes.
The Internal Adjustments menu under System is opened up using the SETUP key.
Adjust All
Starts all internal adjustments for which no external measuring
equipment is needed. The adjustments with external measuring
equipment are described in the Service Manual (supplied).
Remote-control command:
CAL:ALL?
Adjust Synthesis
Performs all adjustments which affect the frequency.
Remote-control command:
CAL:FREQ:MEAS?
Adjust Level
Performs all adjustments which affect the level. The correction values
improve the settling time and the signal quality.
Remote-control command:
CAL:LEV:MEAS?
Adjustment Data
Selects the data used for external level correction.
By default the instrument uses correction data obtained in the factory
before delivery. In addition, customer data can be used for external
level correction. The customer data is obtained using a R&S NRP
power sensor. External level correction is a protected function (see
service manual, chapter 2, "Adjustment").
Remote-control command:
CAL:LEV:EXT:DATA FACT