Description of Individual Diagram Blocks R&S SMC100A
1411.4077.62 2.2 E-1
Description of Individual Diagram Blocks
Available Blocks
Block Function of block Status display in the block Effect of the On checkboxes, or
the MOD ON/OFF and RF
ON/OFF keys
Mod Gen
The internal modulation
signals are configured and
activated.
LF-sweep is configured and
activated.
The internal pulse generator
is configured and activated.
Status of modulation
generator.
Active sweep.
Active pulse generator.
Switches the selected internal LF
generator and the pulse generator
on or off.
Modulation
The modulations are
configured and activated.
Active modulations.
MOD ON/OFF
Switches the active
modulation on or off. The
modulations are activated in
the respective dialog boxes.
RF
RF signal is configured and
activated.
State of RF output.
Activated sweeps.
RF ON/OFF
Switches the RF output on or
off.
Mod Gen block
The internal modulation sources are configured in this block. The LF frequency sweep can be
activated here and an internal LF generator is available as the internal source for the analog
modulations AM, FM and PHiM. The used modulation shape is a sine wave signal.
The internal modulation generator signal is provided at the LF output on the front of the instrument. The
LF output signal and the modulations sources for the analog modulations AM, FM and PhiM can be
selected independently from each other.
A pulse generator provides single and double pulse modulation with selectable pulse widths and
periods. The internal pulse generator signal is provided at the PULSE VIDEO output connector on the
rear panel of the instrument.
The R&S SMC Signal Generator offers three different sweep types (RF frequency sweep, level sweep
and LF frequency sweep) to be activated alternatively. Each type has 6 modes which differ with respect
to the sweep cycle mode (continuous, individual and step-by-step) and triggering mode (automatic,
internal and external). In the Mod Gen block, the LF sweep is configured. Frequency and level sweep
settings are accessed via the RF block.
The status line in the block displays whether the LF generator, the pulse generator and/or a sweep are
active.