R&S SMC100A Alphabetical List of Device-Specific Error Messages
1411.4060.32 9.13 E-1
Adjustment failed (180)
Adjustment could not be executed
Remedy:
The adjustment data have to be generated first by an internal or
external adjustment or to be loaded into the device (see section
Adjustment).
SCPI:
Device-specific error - sets bit 3 in the ESR register
Cannot access the EEPROM (202)
A error occurs when writing or reading a EEPROM.
Example:
The EEPROM is defect.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Cannot access hardware (200)
The data transmission to a module was unsuccessful.
Example:
The module is not installed, not properly installed or missing.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Cannot open file (460)
The selected file can not be opened.
Remedy:
Check the path and file name.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Cannot read file (462)
The file can not be read.
Example:
The file contents are not compatible with the file type.
SCPI:
Device-specific error - sets bit 3 in the ESR register