R&S SMC100A Alphabetical List of SCPI-Error Messages
1411.4060.32 9.11 E-1
Suffix too long (-134)
The suffix contains more than 12 characters.
SCPI:
Command error - sets bit 5 in the ESR register
Syntax error (-102)
The command is invalid.
Example:
The command contains block data the instrument does not
accept.
SCPI:
Command error - sets bit 5 in the ESR register
System error (-310)
This error message suggests an error within the instrument. Please inform the R&S
Service.
SCPI:
Device specific error- sets bit 3 in the ESR register
Too many digits (-124)
The decimal numeric data element contains too many digits.
SCPI:
Command error - sets bit 5 in the ESR register
Too much data (-223)
More data were sent by the host than the signal generator can handle.
SCPI:
Execution error - sets bit 4 in the ESR register
Undefined header (-113)
The sent command header has not been defined.
Example:
Header :*XYZ is undefined for every instrument.
SCPI:
Command error - sets bit 5 in the ESR register