R&S SMC100A RF Phase
1411.4060.32 4.43 E-4
The phase of the output signal can be changed in the Phase Settings section (upper part of the group
menu).
The upper parts concern the frequency settings for the output signal and are described in the section
"RF Frequency", page 4.40
Delta Phase – RF
Phase
Sets the phase of the RF signal. The current phase of the signal is
used as the reference. This function allows, for example, the phase of
the output signal to be synchronized with the phase of a signal from a
second signal generator.
Remote-control command:
SOUR:PHAS 10 DEG
Reset Delta Phase
Display – RF Phase
Resets delta phase value. The set phase is adopted as the new
current phase, i.e. the delta phase value is reset to 0.
Remote-control command:
SOUR:PHAS:REF