Automatic Level Control - ALC R&S SMC100A
1411.4060.32 4.58 E-1
The lower sections provide access to the function User Correction, see section "User Correction",
page 4.59.
State - ALC
Activates/deactivates internal level control.
Off
Sample&Hold
Internal level control is deactivated.
Remote-control command:
SOUR:POW:ALC OFF
SOUR:POW:ALC:OMOD SHOL
Auto
Default state. Level control is automatically adapted
to the operating states.
Remote-control command:
SOUR:POW:ALC AUTO
On
Internal level control is permanently activated.
Remote-control command:
SOUR:POW:ALC ON
Note:
The level is decreased by 30 dB during Sample&Hold measurement!
When pulse modulation is activated, the ALC state of the R&S SMC is automatically changed to
ALC OFF (Sample & Hold). In this state the ALC loop is opened and the output level is not regulated
but the level modulator is set directly. In order to set the correct output level, a sample & hold
measurement is executed after each frequency or level setting. The level is decreased by 30 dB
during Sample&Hold measurement.
Search Once - ALC
Manually activates level control briefly to allow the level to be
calibrated (the Sample&Hold setting must be selected).
Remote-control command:
SOUR:POW:ALC:SONC