R&S SMC100A Sweep Mode
1411.4060.32 4.71 E-4
The sweep range, sweep spacing and dwell time are set in the bottom section.
Execute Single Sweep
– RF Frequency
Sweep
(Mode Single only)
Triggers the sweep manually. A manual sweep can only be triggered if
Mode Single is selected.
Remote-control commands:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC
Reset Sweep – RF
Frequency Sweep
Resets the sweep. The start frequency is set and the next sweep
starts from there.
Remote-control command:
SWE:RES:ALL
Start Freq - RF
Frequency Sweep
Sets the start frequency.
Remote-control command:
SOUR:FREQ:STAR 100MHz
Stop Freq - RF
Frequency Sweep
Sets the stop frequency.
Remote-control command:
SOUR:FREQ:STOP 500MHz
Center Freq - RF
Frequency Sweep
Sets the center frequency.
Remote-control command:
SOUR:FREQ:CENT 300MHz
Span - RF Frequency
Sweep
Sets the span.
Remote-control command:
SOUR:FREQ:SPAN 400MHz
Current Freq - RF
Frequency Sweep
Displays the current frequency.
If Step is set, the frequency for the next frequency step of the sweep
is entered here.
Remote-control command
SOUR:FREQ:MAN 300MHz
Spacing - RF
Frequency Sweep
Selects linear or logarithmic sweep spacing.
Remote-control command:
SOUR:SWE:FREQ:SPAC LIN | LOG