R&S SMC100A LF Generator and LF Output - Mod Gen Block
1411.4060.32 4.93 E-4
The sweep range, sweep spacing and dwell time are set in the bottom section.
State - LF Sweep
Activates LF Sweep mode.
Note:
Activating the LF sweep automatically deactivates the RF
frequency sweep and level sweep.
Remote-control commands:
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:FREQ:MODE CW
Mode - LF Sweep
Selects the Sweep instrument operating mode and Sweep mode.
Auto
Sets an automatic repeated sweep cycle. If a
different sweep mode was activated prior to the
Auto mode, the cycle continues from the current
sweep setting.
The Reset button resets the sweep to the start
frequency.
Remote-control commands:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR AUTO
SOUR:LFO:FREQ:MODE SWE
Single
Sets a single sweep cycle. The sweep is triggered
by the Execute Single Sweep button.
If a different sweep mode was activated prior to the
Single mode, the current sweep is stopped. The
Single sweep always starts at the start frequency.
The Reset button stops the sweep and resets it to
the start frequency.
Remote-control commands:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR SING
SOUR:LFO:FREQ:MODE SWE
SOUR:LFO:SWE:FREQ:EXEC