R&S SMC100A Legend for Front Panel View
1411.4077.62 1.7 E-1
9 Keypad for data entry
DIAGR
Brings the block diagram to the
foreground. Active menus are
minimized.
ESC
CLOSE
Calls the next higher selection level. This opens up
the following functions:
– Closes the active menu if the
cursor is in the top-level menu
(parameter selection).
– Switches between different entry
fields of a menu.
– Quits the editing mode and
restores the previous value. This
function is only available in the
editing mode, i.e. only before an
entry is confirmed or selected with
Enter.
– Cancels queries in message
windows.
– Shifts the entry cursor from the
frequency or level display to the
previously active menu, or to the
previously highlighted block in the
block diagram if no menu is active.
MOD ON/OFF Switches the active modulations on
and off.
RF ON/OFF
Switches the RF signal on and off.
See chapter 3, section
"Setting Parameters",
and chapter 4, section
"RF Signal" and
"Modulations"
10 Standby ON/OFFswitch
The on/off switch switches the instrument from the
standby mode to the operating state provided the
power switch on the rear of the instrument is
switched on.
The yellow LED (right) is on in the standby mode;
the green LED (left) is on when the instrument is
ready for operation.
See chapter 1, section
"Switching On"
CAUTION Danger of shock hazard!
The instrument is still power-supplied while it is in standby mode.