Command Description R&S SMC100A
1411.4060.32 6.8 E-1
CALibration:FMOffset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
The command is a query command and therefore has no *RST value.
Example:
CAL:FMOF?
'starts the adjustments for the FM/PhiM modulator.
Response: 0
'the adjustments have been performed successfully.
*RST value Resolution Options SCPI
- - Device-specific
CALibration:LEVel:EXT:DATA FACTory | CUSTom
Selects the data used for of external level correction.
By default the instrument uses correction data obtained in the factory before delivery. In addition,
customer data can be used for external level correction. The customer data is obtained using a R&S
NRP power sensor. External level correction is a protected function (see service manual, chapter 2,
"Adjustment").
Example:
CAL:LEV:EXT:DATA FACT
'selects the use of the data aquired at the factory for external level correction.
*RST value Resolution Options SCPI
FACTory - Device-specific
CALibration:LEVel[:MEASure]?
Performs all adjustments which affect the level. The aquired correction values improve the settling
time and the signal quality.
The command is a query command and therefore has no *RST value.
Example:
CAL:LEV?
'starts the level adjustment in the mixer range.
Response: 0
'the adjustments have been performed successfully.
*RST value Resolution Options SCPI
- - Device-specific