Command Description R&S SMC100A
1411.4060.32 6.10 E-1
DIAGnostic:BGINfo:CATalog?
The command queries the names of the assemblies available in the instrument.
A complete list of all assemblies is returned (the various entries are separated by commas). The
length of the list is variable and depends on the instrument equipment configuration.
Example:
DIAG:BGIN:CAT?
'
queries the names of the assemblies.
Response: BABO,RFBO,…
*RST value Resolution Options SCPI
- - Device-specific
DIAGnostic:INFO:OTIMe?
The command queries the number of operation hours
Example:
DIAG:INFO:OTIM?
'queries the operation hours.
Response: 100023
'the instrument was operated for 100023 hours up to now.
*RST value Resolution Options SCPI
- - Device-specific
DIAGnostic:INFO:POCount?
The command queries the number of power-on events.
Example:
DIAG:INFO:POC?
'queries the number of power on events.
Response: 123
'the instrument was switched on for 123 times up to now.
*RST value Resolution Options SCPI
- - Device-specific
DIAGnostic:POINt:CATalog?
The command queries the test points available in the instrument. A detailed description of the test
points can be found in chapter 3 , section "Trouble Shooting" of the Service Manual (on CD-ROM,
supplied with the instrument).
Example:
DIAG:POIN:CAT?
'queries the test points available in the instrument.
Response:
'DIAG_SMC_BAB_P7V,DIAG_SMC_BAB_P5V,DIAG_SMC_BAB_P3V3,…'
*RST value Resolution Options SCPI