Command Description R&S SMC100A
1411.4060.32 6.22 E-1
OUTPut - Blank RF Output
OUTPut:BLANk:LIST:STATe ON |OFF
The command activates/deactivates RF output blanking.
The command is a query and therefore has no *RST value.
Example:
":OUTP:BLAN:LIST:STAT ON
'activates the RF output blanking.
*RST value Resolution SCPI
ON - Device-specific
OUTPut - Protection ClearOUTPut<[1]|2>:PROTection:CLEar
The command resets the protective circuit after it has been tripped. The state of the output is again
determined by OUTPut:STATe. .
The output is protected by a protective circuit which deactivates the output in the case of an
externally applied overvoltage This does not change the value of OUTPut:STATe.
Example:
OUTP:PROT:CLE" 'resets the protective circuit for the RF output.
*RST value Resolution Options SCPI
- - Compliant
OUTPut - ProtectionOUTPut:PROTection:TRIPped
The command queries the state of the protective circuit.
The command triggers an event and therefore has no reset value.
Example:
"OUTP:PROT:TRIP?" 'queries the state of the protective circuit for RF output A.
Response:"0" 'the protective circuit has not tripped.
Response:"1" 'the protective circuit has tripped.
*RST value Resolution Options SCPI
- - Compliant
OUTPut[:STATe] ON | OFF
This command activates and deactivates the RF output.
Example:
OUTP OFF
'deactivates the RF output.
*RST value Resolution Options SCPI
- - Compliant