R&S SMC100A Command Description
1411.4060.32 6.25 E-1
SENSe[:POWer]:DISPlay:PERManent:STATe ON | OFFFF
The command switches on and off the permanent indication of the power measurement result in the
upper right corner of the block diagram. The type of sensor, the connector, the measurement source
and - if set - the offset is indicated.
Example:
SENS:POW:DISP:PERM:STAT ON
'the permanent viewer is switched on.
*RST value Resolution SCPI
OFF - -
SENSe[:POWer]:FILTer:LENGth:AUTO?
The command queries the current filter length for auto filter mode (:SENSe:POWer:FILTer:TYPE
AUTO)
The command is a query command and therefore has no *RST value.
Example:
SENS:FILT:TYPE AUTO
'selects auto filter mode for the power sensor connected to the SENSOR
connector.
SENS:FILT:LENG:AUTO?
'queries the automatically set filter length.
Response: 1024
*RST value Resolution SCPI
- Device specific
SENSe[:POWer]:FILTer:LENGth[:USER] 1 65536
The command selects the filter length for user filter mode (SENSe:POWer:FILTer:TYPE USER).
As the filter length works as a multiplier for the time window, a constant filter length results in a
constant measurement time. Values 1 and 2^n are setable.
The time window is fixed to 20 ms.
Example:
SENS:FILT:TYPE USER
'selects user filter mode.
SENS:FILT:LENG 16
'sets a filter length of 16. The resulting measurement time is 640 ms (2x16x20
ms).
*RST value Resolution SCPI
1 - Device specific