EasyManua.ls Logo

Rohde & Schwarz SMC100A - Page 234

Rohde & Schwarz SMC100A
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Description R&S SMC100A
1411.4060.32 6.32 E-1
[SOURce:]AM:SENSitivity?
The command queries the input sensitivity of the MOD EXT input in %/V. The command is only effective
if the external modulation source is selected (SOUR:AM:SOUR EXT). The returned value depends on the
modulation depth setting (SOUR:AM:DEPTh). This value is assigned to the voltage value for full
modulation of the input.
The command is a query command and therefore has no *RST value.
Example:
AM:DEPT 50
'sets a modulation depth of 50 %.
AM:SENS?
'queries the input sensitivity at the MOD EXT input.
Response: 50
'since the voltage value for full modulation is 1V, the resulting sensitivity is
precisely 50%/V.
*RST value Resolution Options SCPI
- - Compliant
[SOURce:]AM:SOURce EXT | INT
The command selects the modulation source for amplitude modulation. INT is the internal
modulation source and is configured in the SOURce:LFOutput subsystem. The external signal is
input at the MOD EXT connector.
Example:
AM:SOUR INT
'selects the internal modulation source.
*RST value Resolution Options SCPI
INTernal - Compliant
[SOURce:]AM:STATe ON | OFF
The command activates/deactivates amplitude modulation.
Example:
AM:STAT ON
'activates AM modulation.
*RST value Resolution Options SCPI
OFF - Compliant

Table of Contents

Other manuals for Rohde & Schwarz SMC100A

Related product manuals