R&S SMC100A Command Description
1411.4060.32 6.71 E-1
[SOURce:]PULM:TRIGger:EXTernal:GATe:POLarity NORMal | INVerted
Selects the polarity of the Gate signal. The signal is supplied via the PULSE EXT connector.
Example:
PULM:TRIG:EXT:GAT:POL NORM
'The pulse signal is generated while the gate signal is high.
*RST value Resolution Options SCPI
NORMal - Device-specific
[SOURce:]PULM:TRIGger:MODE AUTO | EXTernal | EGATe
The command selects the trigger mode for pulse modulation.
Parameters: AUTO
The pulse modulation is generated continuously.
EXTernal
The pulse modulation is triggered by an external trigger event. The trigger
signal is supplied via the PULSE EXT connector.
EGATe
The pulse modulation is gated by an external gate signal. The signal is
supplied via the PULSE EXT connector.
Example:
PULM:TRIG:MODE EXT
'selects triggering by an external trigger event. The trigger signal is supplied
via the PULSE EXT connector.
*RST value Resolution Options SCPI
AUTO - Compliant
[SOURce:]PULM:WIDTh 20ns .. 1s
The command sets the width of the generated pulse. The width determines the pulse length. The
pulse width must be at least 20 ns less than the set pulse period.
Example:
PULM:WIDT 33 us
'sets a width of 33 us for the pulse.
*RST value Resolution Options SCPI
1ms 20 ns Device-specific