R&S SMC100A Command Description
1411.4060.32 6.85 E-1
STATus:PRESet
The command resets the status registers. All PTRansition parts are set to FFFFH (32767), i.e. all
transitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a transition from 1 to 0 in
a CONDition bit is not detected. The ENABle parts of STATus:OPERation and
STATus:QUEStionable are set to 0, i.e. all events in these registers are not passed on.
The command triggers an event and therefore has no query form and no *RST value.
Example:
STAT:PRES
'resets the status registers.
*RST value Resolution Options SCPI
- - Compliant
STATus:QUEStionable:EVENt?
The command queries the content of the EVENt part of the STATus:QUEStionable register. This part
contains information on the actions performed in the instrument since the last readout. The content
of the EVENt part is deleted after being read out.
Example:
STAT:OPER:EVEN?
'queries the Status:Questionable:Event register.
*RST value Resolution Options SCPI
- - Compliant
STATus:QUEStionable:CONDition?
The command queries the content of the CONDition part of the STATus:QUEStionable register. This
part contains information on the action currently being performed in the instrument. The content is
not deleted after being read out since it indicates the current hardware status.
Example:
STAT:OPER:COND?
'queries the Status:Questionable:Condition register.
*RST value Resolution Options SCPI
- - Compliant
STATus:QUEStionable:PTRansition
The command sets the bits of the PTRansition part of the STATus:QUEStionable register. If a bit is
set, a transition from 0 to 1 in the condition part causes an entry to be made in the EVENt part of the
register.
Example:
STAT:OPER:PTR 32767
'all transitions from 0 to 1 in the condition part of the Status:Questionable
register cause an entry to be made in the EVENt part.
*RST value Resolution Options SCPI
- - Compliant