Command Description R&S SMC100A
1411.4060.32 6.94 E-1
SYSTem:DISPlay:UPDate ON | OFF
The command switches the update of the display on/off. A switchover from remote control to manual
control always sets the status of the update of the display to ON.
Example:
SYST:DISP:UPD OFF
'switches update of displayed parameter values off.
*RST value Resolution Options SCPI
- - Device-specific
SYSTem:ERRor:ALL?
The command queries all entries in the error queue and then deletes them. Positive error numbers
denote device-specific errors, and negative error numbers denote error messages defined by SCPI
(see section "Error Messages"). If the error queue is empty, 0 ("No error") is returned.
The command is a query command and therefore has no *RST value.
Example:
SYST:ERR:ALL?
'queries all entries in the error queue.
Response: "0, 'no error'"
'no errors have occurred since the error queue was last read out.
*RST value Resolution Options SCPI
- - Compliant
SYSTem:FPReset
The command triggers an instrument reset to the original state of delivery. This includes also the
parameters, that are unchanged when executing a regular Preset with the *RST command, e.g.
IEC/IEEEE bus address. However, password settings and all settings protected by these password
are not changed, e.g. disabled USB connections, LAN settings and reference oscillator setting.
The command triggers an event and therefore has no *RST value and no query form.
Example:
SYST:FPR
'all instrument settings (also those that are not currently active) are reset to the
factory values.
*RST value Resolution Options SCPI
- - Device-specific