7-46
TSL-710 Operation Manual
6. Command error
Errors issued in Command set 1 are stored as error messages in the error
queue and can be read out with the ”:SYSTem:ERRor?” command. The list is
as shown below.
Code Error message
0 No error
-102
Syntax error
-103
Invalid separator
-108
Parameter not allowed
-109
Missing parameter
-113
Undened header
-148
Character data not allowed
-200
Execution error
-222
Data out of range
-410
Query INTERRUPTED
Command Error List (Command set 1 Group)
[:SOURce]:AM:STATe?
Syntax [:SOURce]:AM:STATe?
Description  Reads out status of amplitude modulation of the laser output.
Parameter None
Response 0:Disable
1:Enable
Example :AM:STAT?
1
Indicates that the amplitude modulation is enabled.